Understanding OSHA’s Calibration Standards
The Occupational Safety and Health Administration (OSHA) mandates daily calibration checks as part of its OSHA calibration requirements before conducting audiometric tests on employees. This requirement underscores OSHA’s commitment to ensuring the safety and health of workers, particularly in the realm of hearing protection. For these tests to yield accurate and reliable results, the equipment must be calibrated correctly.
There are two primary methods to ensure this calibration:
- Using a reliable test subject with standard hearing to compare their responses to the audiometer’s known calibration levels.
Daily Functional Check
The purpose of the Daily Functional Check is to ensure that the cable connections are interruption-free and that there is no abnormality (static, volume changes) in the sound output of the testing system. For this test, ANSI recommends using a 1000 Hz tone at 70 dB(HL) and requests that the test administrator put on the headphones and present these tones one at a time in each ear, all while putting slight pressure on the headphone jack connectors, and wiggling the cables to listen for any abnormalities. The tone presentation time should be approximately 20 seconds on each ear to allow for adequate manipulation of the jacks and cables. Should abnormalities be discovered, these should be addressed prior to continuing with testing. The Daily Functional Check is a static one-time verification each testing day, and the results should be recorded to help with the possible future investigation of an emerging equipment issue or in the case of an audit request for
equipment verification practices.
Daily Biological Verification
The Biological Verification aims to ensure that there hasn’t been significant sound output drift on a per-frequency basis since the last annual calibration. In order to set up this test, it is recommended that you create a distinct result file for this purpose for each individual who may administer workplace hearing testing. For example, creating a file with the patient name ‘Biological Verification A.A.’ under which every test can be saved and compared to baseline. Only individuals with stable hearing levels should perform Biological Verification.
After the first test for an individual has been marked as a ‘baseline,’ each test after that can be compared frequency-by-frequency for deviations of more than 10 dB(HL). Should there be deviations of more than 10 dB(HL), the system will require additional attention, and we would recommend contacting your Professional Supervisor or Hearing Conservation Program Administrator for recommended next steps.